Details of Equipment: |
|
Product Type: |
Short Range Radio Device |
Brand Name: |
Medtronic |
Model Number: |
Serena CRT-P MRI SureScan |
NCA Equipment No.: |
ZRO-M8-7E3-202 |
Product Name: |
Serena CRT-P MRI SureScan |
Intended Use: |
Active Implantable Medical Device |
Additional Details for Radio Equipment: |
Frequency Range: |
175 kHz, 2402-2480 MHz |
RF Output Power Radiation: |
175 kHz: 3.1 dBuA/m at 10m; 2402-2480 MHz: -6.19 dBm EIRP |
RF Output Power Conducted: |
NA |
RF Channel Spacing: |
175 kHz: single channel, <50 kHz; 2402-2480 MHz: 40 channels, <2 MHz |
RF Output Impedance: |
50 Ohm |
Type Of Modulation: |
175 kHz: PSK uplink, FSK downlink modulation; 2402-2480 MHz: GFSK |
Bandwidth: |
175 kHz: <50 kHz; 2402-2480 MHz: <2 MHz |
Software Version: |
NA |
Antenna Type: |
175 kHz: Integral loop antenna; 2402-2480 MHz: monopole antenna |
Antenna Gain: |
175 kHz: N/A; 2402-2480 MHz: -24.02 dBi |
Technical Variants: |
NA |
Details of Certificate of Compliance: |
Issuing Body: |
FCC, ISED, CE |
Issuing Date: |
8/10/2017 |
Validity: |
no expiration |
Standards: |
EMC: |
EN 301 489-31
EN 301 489-17 |
Radio: |
EN 302 195
EN 300 328 |
Health And Safety: |
EN 45502-1
EN 45502-2-2 |
Manufacture's Information: |
Name Of Manufacturer: |
Medtronic, Inc. |
Address Of Manufacturer: |
710 Medtronic Parkway, Fridley, MN-55432, USA |